2-2 | Thermox Series 2000 WDG Insitu Analyzer
Sensor
Principle of
Operation
Direct sensing insitu probe suitable for applications where the flue gas
temperature does not exceed 1250°F (677°C) and where combustibles
measurement is not required. Zirconium oxide for net oxygen mea-
surement.
Output Range
From 0-1% to 0-100%
Accuracy
± 1% of measured value or ± 0.05% oxygen, whichever is greater
Response
63% of a process step change < 20 secs.
Drift
< 0.1% of cell output per month (< 0.005% O
2
per month with 2% O
2
applied)
Linearity
± 1% of output span
Repeatability
± 1% of output span
Max. Flue Gas Temp / Probe Type Material
1250°F (677°C)
310 SS
Sample Pressure
± 10 inch water gauge
Environment
Ambient Temp.:
-5°F to 160°F (-20°C to 70°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity:
0% to 90%, non-condensing
IEC Installation
Overvoltage Category II
IEC Pollution Degree 2
Max Altitude::
5000 meters
Enclosure
UL Type (NEAM 4X) standard. Optional explosion-proof and purged versions available. Purged
version for sub-atmospheric process pressure only.
Power Requirements
115 VAC, ± 10%, 47-63 Hz, 150 VA max. 230 VAC, ± 10%, 47-63 Hz, 150 VA max.
Calibration Gas Requirements
Use calibration gases @ 10 PSIG, 3 to 6 SCFH (0.70 kg/cm
2
, 1.41 to 2.82 L/min.)
O
2
Span Gas: Air or from 1.0 to 100% O
2
, balance N
2
O
2
Zero Gas: 2% or from 0.1 to 10% O
2
, balance N
2
System Compliance
EMC Compliance:
Safety Compliance:
2004/108/EC
73/23/EEC
NOTES
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...