5-6 | Thermox Series 2000 WDG Insitu Analyzer
2. Select the
Auto Calibrate
or
Auto Verify
option. The system will begin
the automatic calibration, and will notify you how long the span gas
will be applied to the sensor from the RCU:
Span Gas xx:xx
System Calibrating
Then the following message will appear on your display:
Zero Gas xx:xx
System Calibrating
During this time, the RCU switched the solenoid valves and introduced
the zero gas to the sensor.
When specified, the system will display the recovery time remaining:
Recovery xx:xx
The system returns to monitoring process readings and the calibration
or verify operation is complete. You can view the results of the last cali-
bration or verification using the
Cal/Verify Data
menu option from the
Calibrate
key menu.
Remote Calibrate/Remote Verify
To initiate a remote calibration or verification, do the following:
1. Enter the calibration gas values to match the span and zero calibration
gas cylinders you intend to use to calibrate/verify your system. These
values are referred to in this section as
setpoint values
.
2. Select
Initiate Cal
or
Initiate Verify
from the
Calibrate
key menu, de
-
pending on whether you want to perform a calibration or verification.
3. Select
Remote Calibrate
or
Remote Verify
.
The following appears on your display:
O2 xx.x % Set xx.x %
Press ENTER to Span
4. Press
Enter
when the O
2
reading on the far left of the first display line
stabilizes. The reading on the far right of the first display line shows the
setpoint value you entered using the
Cal Gas Values
menu option. This
setpoint value should match the cal gas cylinder value.
If your O
2
reading on the far left of the display has stabilized, yet is not near the setpoint
value on the far right of the display, you should check the cal gas value you entered and
ensure that this value matches the cal gas cylinder for that gas in the RCU.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...