Calibration | 5-17
Auto Cal is ENABLED
or
Auto Cal is DISABLED
and will then exit this menu option.
Set Cal Timers - Cal Gas Duration
Define how long the RCU should apply the span and zero calibration
gases to the sensor during an automatic calibration or verification. You can
also set a calibration gas duration time to 0 if you don’t want to include it
in your automatic calibration sequence.
To find out how long it takes for your calibration gases to reach the sen-
sor and stabilize, select the
Inject Cal Gas
menu option and time the gas
stabilization period. Then use this time frame as your calibration gas dura-
tion (we suggest that you also add a buffer to this time frame to be certain
that the gas has stabilized).
1. From the
Set Cal Timers
menu option, select
Cal Gas Duration
.
You will be prompted to enter the new span gas duration time:
O2 Span Time = xx:yy
Enter new time: Mn:Sc
2. Enter the new span gas duration time, then press
Enter
.
Mn
equals
minutes and
Sc
= seconds. As you type the new span gas duration
time, it overwrites the Mn:Sc text. If you want to use the currently
displayed time, press
Enter
without entering a new value.
At this point, you will be prompted to enter the zero gas duration time:
O2 Zero Time = xx:yy
Enter new time: Mn:Sc
3. Enter the new zero gas duration time, then press
Enter
.
Mn
equals
minutes and
Sc
= seconds. As you type the new zero gas duration
time, it overwrites the Mn:Sc text. If you want to use the currently
displayed time, press
Enter
without entering a new value.
Set Cal Timers - Recovery Duration
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...