Commands in DSR Application Subsystems
Optical Input Commands
264
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:INPut:OPTic[:STATe][:VALue]
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):INPut:OPTic:[STATe]:
[VALue] <OFF|ON>
:<Handle>:SGENeral:PDATa(*):INPut:OPTic:[STATe]:[VALue] <OFF|ON>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):INPut:OPTic:[STATe]:[VALue]
<OFF|ON>
Parameters
<OFF|ON>
Sets the state of the optical channel(s) when in optical
mode. The electrical analyzer is automatically enabled/disabled.
Examples
:_test:CGR1:MOD1:CONN1:INP:OPT:STAT:VAL OFF
:_test:CGR1:MOD1:CONN1:INP:OPT OFF
:INPut:OPTic[:STATe][:VALue]?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):INPut:OPTic:[STATe]:
[VALue]? <NFORce|FORCe>
:<Handle>:SGENeral:PDATa(*):INPut:OPTic:[STATe]:[VALue]?
<NFORce|FORCe>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):INPut:OPTic:[STATe]:[VALue]?
<NFORce|FORCe>
Return Value
The state of the optical channel(s) when in optical mode.
Parameters
<NFORce|FORCe>
Defines how the query is executed:
• NFORce
Default value. Only the firmware is queried for the current state.
• FORCe
When FORCe is entered, the firmware checks the current condition
in the hardware. This is required to make sure that the optical
channel's electrical interface is really switched on.
N O T E
The
FORCe
query is slower than
NFORce
.
Examples
:_test:CGR1:MOD1:CONN1:INP:OPT:STAT:VAL? FORC
:_test:CGR1:MOD1:CONN1:INP:OPT?
might return:
ON