Commands in DSR Application Subsystems
:SGENeral:GLOBal Subsystem
142
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:SGENeral:GLOBal Subsystem
Syntax
:<Handle>:SGENeral:GLOBal
This subsystem provides commands to modify system parameters.
These commands affect the entire system including any frames
connected via MXI.
In addition to the top-level commands, this subsystem contains
further subsystems:
•
“:SGENeral:GLOBal:CONFigure Subsystem” on page 151
•
“:SGENeral:GLOBal:INITiate:CONTinuous Subsystem” on page 154
•
“:SGENeral:GLOBal:SYSTem Subsystem” on page 155
•
“:SGENeral:GLOBal:SEQuence Subsystem” on page 156
•
“:SGENeral:GLOBal:TRIGger Subsystem” on page 173
•
“:SGENeral:GLOBal:ARM Subsystem” on page 179
:<Handle>:SGENeral:GLOBal:CONNect
Syntax
:<Handle>:SGENeral:GLOBal:CONNect <ON | OFF>
Description
Connects or disconnects all enabled connectors of the Agilent 81250
System at the same time.
Example
:_test:SGEN:GLOB:CONN OFF
:<Handle>:SGENeral:GLOBal:CONNect?
Syntax
:<Handle>:SGENeral:GLOBal:CONNect?
Return Value
Returns the actual connection status of the enabled connectors.
Example
:_test:SGEN:GLOB:CONN?
might return:
OFF
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com