:EDIT:SEGMent(*) Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
103
:<Handle>:EDIT:SEGMent(*):CLOSe
Syntax
:<Handle>:EDIT:SEGMent(*):CLOSe
Parameters
(*)
Inspector number of the segment.
Description
Closes the segment editor specified by the segment inspector index
number without saving any changes made to the current segment
name.
Example
:_test:edit:segm3:clos
:<Handle>:EDIT:SEGMent(*):PATTern:DATA
Syntax
:<Handle>:EDIT:SEGMent(*):PATTern:DATA
<trace1>,<vector1>,<trace2>,<vector2>,<stream>
Parameters
<trace1>
Number of start trace.
<vector1>
Number of start vector.
<trace2>
Number of end trace.
<vector2>
Number of end vector.
<stream>
Data stream to be set in the segment (hexadecimal
string).
(*)
Inspector number of the segment.
Description
Sets data in a bounded rectangle in the segment.
New data can only be set for database segments. Attempting to set new
data in an analyzer segment will cause an error.
The input data stream (<stream>) is a string of hexadecimal
characters. The length of this string must be at least as many
hexadecimal characters as needed to provide all the information
required for the bounded rectangle. The format of this string is the
same as explained for
“:<Handle>:EDIT:SEGMent(*):PATTern:DATA?”
on page 104
.
If more characters are provided than required, they will be ignored.
Also the trailing bits for the last character in every trace will be
ignored.
Example
:_test:EDIT:SEGM3:PATT:DATA 0,0,3,2,"FFFF"