Commands in DSR Application Subsystems
DeMUX Rewiring Parameter Commands
222
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:
<
Handle
>
:SGENeral:GLOBal:
SYNChronization:RCH?
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronizat:RCHeck?
Return Value
Returns the current rewire verify mode (on or off).
If verify rewiring is switched on, the system performs additional
inspections and complains about possible errors.
If verify rewiring is switched off, the system tries to ignore noncritical
problems to carry out a measurement.
Example
:_test:SGEN:GLOB:SYNC:RCH?
might return:
ON or OFF
:<Handle>:SGENeral:PDATa:DMUX:STAGe
Syntax
:<Handle>:SGENeral:PDATa*:DMUX:STAGe*[:VALUE]
Description
Specifies the number of stages of the DeMUX module. To set this value,
you need to know the architecture of the demultiplexer.
Parameters
*.
Number of stages of the DeMUX module.
Example
For a demultiplexer with the following architecture:
:_test:SGEN:PDATA:DMUX:STAG 3
Stage
3
Stage
2
Stage
1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com