Commands in DSR Application Subsystems
[:CGRoup(*)]:MODule(*):CONNector(*) Subsystem
132
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Related Commands
•
“:<Handle>:SGENeral:GLOBal:CALibration:SELF DELay” on
page 143
Example
:_test:CGR1:MOD1:CONN:CAL:ZDEL 1.0e-9
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):CALibration:ZDELay?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):CALibration:ZDELay?
Return Value
Returns the current zero delay for the specified connector.
Example
:_test:cgr1:mod1:conn1:cal:zdel?
might return:
1.000000E-009
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):SPECies[:VALue]?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):SPECies?
Return Value
Returns a comma-separated list of the species states that are
supported by the connector. This can be either ELECtric or OPTic.
Example
:_test:cgr1:mod1:conn1:spec?
might return:
ELEC,OPT
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):SPECies:STATe
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):SPECies:STATe
<ELEC|OPT>
Parameters
<ELEC|OPT>
Switches the connector into either electrical (ELEC)
or optical (OPT) mode.
N O T E
This command can only be executed when the connector is not
connected to any port.
Example
:_test:cgr1:mod1:conn1:spec:stat elec