Commands in DSR Application Subsystems
Input Parameter Commands
252
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:INPut:IMPedance[:INTernal]
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:IMPedance[:INTernal] <Impedance>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:IMPedance[:INTernal]
<Impedance>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:IMPedance[:
INTernal] <Impedance>
Parameters
<Impedance>
Internal termination impedance value.
Specifies the internal termination impedance of the analyzer input
connector. To set 10 kOhm impedance, use -1 in the command. For
50 Ohm impedance, use 50 in the command.
Example
:_test:SGEN:PDAT1:INPut:IMP -1
:_test:SGEN1:PDAT1:TERM1:INPut:IMP -1
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:IMP -1
:INPut:IMPedance[:INTernal]?
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:IMPedance[:INTernal]?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:IMPedance[:INTernal]?
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:IMPedance[:
INTernal]?
Return Value
Returns the current internal impedance of the specified analyzer input
connector. A negative value relates to the HiZ internal impedance.
Example
:_test:SGEN:PDAT1:INPut:IMP?
:_test:SGEN1:PDAT1:TERM1:INPut:IMP?
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:IMP?
might return:
-1.000
:INPut:SEQuence:[LAYer:SOURce:]CLOCk
Syntax
:<Handle>:SGeneral:PData:INPut:SEQuence:[LAYer:SOURce:]CLOCk
<INTernal | CDR>
:<Handle>:SGeneral:PData:TERMinal:INPut:SEQuence:[LAYer:SOURce:
]CLOCk <INTernal | CDR>