Commands in DSR Application Subsystems
Output Parameter Commands
272
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:OUTPut:TVOLtage
Syntax
:<Handle>:SGENeral:
PDATa(*):OUTPut:
TVOLtage <TerminationVoltage>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):OUTPut:
TVOLtage
<TerminationVoltage>
:<Handle>:SGENeral:
PPULse(*):OUTPut:
TVOLtage <TerminationVoltage>
:<Handle>:SGENeral:
PPULse(*):TERMinal(*):OUTPut:
TVOLtage
<TerminationVoltage>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):OUTPut:
TVOLtage
<TerminationVoltage>
Parameters
<TerminationVoltage>
Termination voltage (<NRf>).
Sets the external termination voltage, default is 0 volts = ground.
Valid range is –2.0 to +3.0 volts.
Example
:_test:SGEN:PDAT1:OUTP:TVOLT -1
:_test:SGEN:PDAT1:TERM1:OUTP:TVOL 2.0
:_test:SGEN:PPUL1:OUTP:TVOLT -1
:_test:SGEN:PPUL1:TERM1:OUTP:TVOL 2.0
:_test:CGROUP1:MODULE2:CONNECTOR4:OUTPUT:TVOLTAGE -1
:OUTPut:TVOLtage?
Syntax
:<Handle>:SGENeral:
PDATa(*):OUTPut:
TVOLtage?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):OUTPut:
TVOLtage?
:<Handle>:SGENeral:
PPULse(*):OUTPut:
TVOLtage?
:<Handle>:SGENeral:
PPULse(*):TERMinal(*):OUTPut:
TVOLtage?
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):OUTPut:
TVOLtage?
Return Value
Returns the currently set value for the external Termination VOLtage.
Example
:_test:SGEN:PDAT1:OUTP:TVOLT?
:_test:SGEN:PDAT1:TERM1:OUTP:TVOL?
:_test:SGEN:PPUL1:OUTP:TVOLT?
:_test:SGEN:PPUL1:TERM1:OUTP:TVOL?
:_test:CGROUP1:MODULE2:CONNECTOR4:OUTPUT:TVOLTAGE?
might return:
-1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com