Commands in DSR Application Subsystems
:SGENeral:GLOBal:ARM Subsystem
182
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:<Handle>:SGENeral:GLOBal:ARM[:
SEQuence][:LAYer]:THReshold
Syntax
:<Handle>:SGENeral:GLOBal:ARM[:SEQuence][:LAYer]:THReshold <NRf>
This command specifies the threshold of the external input signal
which is used in the
GATed | STARtsignal | STOPsignal
modes. The
range is from -1.40 V to +3.70 V.
Example
:_test:SGEN:GLOB:ARM:THR 1
:<Handle>:SGENeral:GLOBal:ARM[:
SEQuence][:LAYer]:THReshold?
Syntax
:<Handle>:SGENeral:GLOBal:ARM[:SEQuence][:LAYer]:THReshold?
Returns the threshold for the external input signal used in the gate or
start/stop mode.
Example
:_test:SGEN:GLOB:ARM:THR?
might return:
1.000
:<Handle>:SGENeral:GLOBal:ARM[:
SEQuence][:LAYer]:TVOLtage
Syntax
:<Handle>:SGENeral:GLOBal:ARM[:SEQuence][:LAYer]:TVOLtage <NRf>
This command specifies the termination voltage of the external input
signal which is used in the GATed | STARtsignal | STOPsignal modes.
The range is from -2.10 V to +3.30 V.
Example
:_test:SGEN:GLOB:ARM:TVOL -2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com