Output Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
269
:OUTPut[:STATe]?
Syntax
:<Handle>:SGENeral:
PDATa(*):OUTPut
[:STATe]? <NFORce|FORCe>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):OUTPut:
[STATe]?
<NFORce|FORCe>
:<Handle>:SGENeral:
PPULse(*):OUTPut
[:STATe]? <NFORce|FORCe>
:<Handle>:SGENeral:
PPULse(*):TERMinal(*):OUTPut:
[STATe]?
<NFORce|FORCe>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):OUTPut:
[STATe]?
<NFORce|FORCe>
Parameters
<NFORce|FORCe>
Defines how the query is executed:
• NFORce
Default value. Only the firmware is queried for the current state.
• FORCe
When FORCe is entered, the firmware checks the current condition
in the hardware. This is required to make sure that the output is
really switched on, or to check if the protection circuit within the
output has turned off the channel to protect it from damage.
N O T E
The
FORCe
call is slower than
NFORce
.
Return Value
Returns the current connection state for the specified output
connectors.
:_test:SGEN:PDAT1:OUTP? NFOR
:_test:SGEN:PDAT1:TERM1:OUTP:STAT? FORC
:_test:SGEN:PPUL1:OUTP? NFOR
:_test:SGEN:PPUL1:TERM1:OUTP:STAT? FORC
:_test:MOD2:CONN1:OUTP?
might return:
ON