:SGENeral:GLOBal:SYNChronization Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
171
:<Handle>:SGENeral:GLOBal:
SYNChronization:APSelect?
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronization:APSelect?
Return Value
Returns whether the Automatic Polarity Select feature is enabled or
disabled. Use
“:INPut:POLarity?” on page 246
to query the signal’s
polarity.
Example
:_test:SGEN:GLOB:SYNC:APS?
might return:
TRUE
:<Handle>:SGENeral:GLOBal:
SYNChronization:APAlignment
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronization:APAlignment <FALSe | TRUE>
Parameters
Specifies whether a phase optimization is done after an automatic bit
synchronization.
TRUE
The phase is automatically optimized as specified by the
SGEN:GLOB:SYNC:BERT and SGEN:GLOB:SYNC:PACC commands.
This is the reset value.
FALSe
No automatic phase alignment.
Note that this value will only be used if SGEN:GLOB:SYNC:USED?
returns TRUE and SGEN:GLOB:SYNC:SMOD? returns BSYN.
Example
SGEN:GLOB:SYNC:APA TRUE