Amplitude/Offset Correction Factor Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
237
Example
:_test:SGEN:PDAT1:CAL:ACF?
might return:
7.5E–1
:CALibration[:VOLTage]:HOLD
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CALibration[:VOLTage]:HOLD
<IND| LOCK>
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CLOCk:CALibration[:
VOLTage]:HOLD <IND| LOCK>
:<Handle>:SGENeral:PDATa(*):CALibration[:VOLTage]:HOLD <IND| LOCK>
:<Handle>:SGENeral:PDATa(*):CLOCk:CALibration[:VOLTage]:HOLD <IND|
LOCK>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CALibration[:VOLTage]:HOLD
<IND| LOCK>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CLOCk:CALibration[:VOLTage]:
HOLD <IND| LOCK>
Parameters
<IND|LOCK>
Defines whether the offset and amplitude correction
factors are independent (IND) or locked (LOCK). See
“Amplitude/Offset Correction Factor Commands” on page 234
for
details of this command.
Example
:_test:sgen:pdat1:cal:hold LOCK
:CALibration[:VOLTage]:HOLD?
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CALibration[:VOLTage]:
HOLD?
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CLOCk:CALibration[:
VOLTage]:HOLD?
:<Handle>:SGENeral:PDATa(*):CALibration[:VOLTage]:HOLD?
:<Handle>:SGENeral:PDATa(*):CLOCk:CALibration[:VOLTage]:HOLD?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CALibration[:VOLTage]:HOLD?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CLOCk:CALibration[:VOLTage]:
HOLD?
Return Value
Returns the current correction factor locking mode. See
“Amplitude/Offset Correction Factor Commands” on page 234
for
details of this command.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com