Commands in DSR Application Subsystems
Connector Administration Commands
310
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:TERMinal(*):[TO]
Syntax
:<Handle>:SGENeral:CONNect:
PDATa(*):TERMinal(*):
[TO] <@ChannelList>
:<Handle>:SGENeral:CONNect:
PPULse(*):TERMinal(*):
[TO] <@ChannelList>
Parameters
<@ChannelList>
List of channels.
Connects the specified terminal to the specified channel or channels.
If the number of connectors exceeds the number of terminals defined
in the specified port, the connection will be continued to the next port.
In other words, the suffix of the port and the suffix of the terminal are
used as a start value. Only connectors of the same type can be
combined to a port.
The syntax of the channel list is as follows:
Syntax Examples:
• (@0102004) specifies connector 4 in clock group 1, module 2
• (@0102001:0102004,0103001) specifies connectors 1 to 4 of module
2 and connector 1 of module 3, both in clockgroup 1
Example
:_test:SGEN:CONN:PDAT1:TERM2:TO (@0102001)
:_test:SGEN:CONN:PPUL1:TERM2:TO (@0102001)
<ChannelListExpr> ::= “(@”<ChannelList>“)”
<ChannelList>
::= <ChannelListElem> [“,” <ChannelListElem>]
<ChannelListElem> ::= <Connector> | (<Connector> “:” <Connector>)
<Connector>
::= <ClockGroupNr><ModuleNr><ConnectorNr>
<ClockGroupNr>
::= <Digit><Digit>
<ModuleNr>
::= <Digit><Digit>
<ConnectorNr>
::= <Digit><Digit><Digit>
<Digit>
::= “0” | “1” | “2” | “3” | “4” | “5” | “6” | “7” | “8” |
“9”