Clock Output Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
297
• Via a resistor to the complement output. Optimum signal
performance is achieved with 100 Ohms.
Example
:_test:CGR:MOD2:CONN:CLOC:OUTP:TCON VOLT
:CLOCk:OUTPut:TCONfig?
Syntax
CGRoup(*):MODule(*):CONNector(*)
SGENeral:PDATa(*)
SGENeral:PDATa(*):TERMinal(*)
Return Value
Returns the termination module for the N4872/74’s clock output.
Example
:_test:CGR:MOD2:CONN:CLOC:OUTP:TCON?
might return:
VOLT
:CLOCk:OUTPut:TVOLtage
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CLOCk:OUTPut:TVOLtage
<NRf>
:<Handle>:SGENeral:PDATa(*):CLOCk:OUTPut:TVOLtage <NRf>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CLOCk:OUTPut:TVOLtage <NRf>
Parameters
<NRf>
Sets the external termination voltage of the N4872/74’s clock
output, default is 0 volts (ground). Valid range is –2.0 to +3.0 volts.
Example
:_test:CGR:MOD2:CONN:CLOC:OUTP:TVOL 1.0
100
OUT
~OUT