Commands in DSR Application Subsystems
Timing Parameter Commands
190
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:PULSe:CROSsing?
Syntax
:<Handle>:SGENeral:PDATa(*):[:SOURce]:PULSe:CROSsing?
:<Handle>:SGENeral:PDATa(*):TERMinal(*)[:SOURce]:PULSe:CROSsing?
:<Handle>:SGENeral:PPULse(*):[:SOURce]:PULSe:CROSsing?
:<Handle>:SGENeral:PPULse(*):TERMinal(*)[:SOURce]:PULSe:CROSsing?
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*)[:SOURce]:PULSe:CROSsing?
Return Value
Returns the current cross point. The unit for the cross point is in
percentage of signal amplitude. The formula for the cross point is:
Cross Point = 100 * ( (V
Cross
– V
High
) / (V
High
– V
Low
) )
Example
:_test:SGEN:PDAT1:PULS:CROS?
:_test:SGEN:PDAT1:TERM1:SOUR:PULS:CROS?
:_test:SGEN:PPUL1:PULS:CROS?
:_test:SGEN:PPUL1:TERM1:SOUR:PULS:CROS?
:_test:MODULE2:CONNECTOR3:PULSE:CROSSING?
might return:
50.0