Commands in DSR Application Subsystems
:SGENeral:GLOBal:TRIGger Subsystem
176
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Example
:_test:SGEN:GLOB:TRIG:CLOC:MEAS:DIR
or
:_test:SGEN:GLOB:TRIG:CLOC:MEAS
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:MULTiplier
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:MULTiplier
<Multiplier>
Parameter
<Multiplier>
Sets the clock multiplication factor in a range of 1 ...
256.
This command works only when the Agilent 81250 has been stopped.
Example
:_test:SGEN:GLOB:TRIG:CLOC:MULT 10
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:MULTiplier?
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:
MULTiplier?
Return Value
Returns the current clock multiplication factor.
This command works only when the Agilent 81250 has been stopped.
Example
:_test:SGEN:GLOB:TRIG:CLOC:MULT?
might return:
10
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:DIVider
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:DIVider
<DIVider>
Parameters
<DIVider>
The clock divider.
Range/Value Coupling
Period, frequency, measured value.
Reset Value
1