Commands in DSR Application Subsystems
[:CGRoup(*)][:SOURce]:TRIGger Subsystem
138
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:<Handle>[:CGRoup(*)][:SOURce]:TRIGger:
TVOLtage?
Syntax
:<Handle>[:CGRoup(*)][:SOURce]:TRIGger:TVOLtage?
Return Value
Returns the current termination voltage of the trigger output
connector.
Example
:_test:CGROUP1:SOURCE:TRIGGER:TVOLTAGE?
might return:
-2.000
:<Handle>[:CGRoup(*)][:SOURce]:TRIGger:
IMPedance:EXTernal
Syntax
:<Handle>[:CGRoup(*)][:SOURce]:TRIGger:IMPedance:EXTernal <NRf>
Description
The external termination impedance can be specified by this
command. Any negative value is interpreted as into “open”.
Example
:_test:CGR1:TRIG:IMP 500
:<Handle>[:CGRoup(*)][:SOURce]:TRIGger:
IMPedance:EXTernal?
Syntax
:<Handle>[:CGRoup(*)][:SOURce]:TRIGger:IMPedance:EXTernal?
Return Value
Returns the current programmed external termination impedance
(load impedance).
Example
:_test:CGR1:TRIG:IMP?
might return:
5.002
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com