Commands in DSR Application Subsystems
Auxiliary Out Commands
298
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:CLOCk:OUTPut:TVOLtage?
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CLOCk:OUTPut:TVOLtage?
:<Handle>:SGENeral:PDATa(*):CLOCk:OUTPut:TVOLtage?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CLOCk:OUTPut:TVOLtage?
Return Value
Returns the current termination voltage of the N4872/74’s clock
output.
Example
:_test:CGR:MOD2:CONN:CLOC:OUTP:TVOL?
might return:
3.0E+0
Auxiliary Out Commands
The Auxiliary Out command controls the output at the AUX out of an
N4873A/75A module.
:AOUT[:MONitor]
Syntax
CGRoup(*):MODule(*):CONNector(*):AOUT[:MONitor] <DME | CLOC>
SGENeral:PDATa(*):AOUT[:MONitor] <DME | CLOC>
SGENeral:PDATa(*):TERMinal(*):AOUT[:MONitor] <DME | CLOC>
Parameters
<DME | CLOC>
Defines the signal at the AUX output of a
N4873A/75A module. DMEasurement selects the data stream and
CLOCk selects the (recovered) clock.
Example
:_test:sgen:pdat:aout DME
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com