Optical Output Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
287
:OUTPut:OPTic:POWer:MAMPlitude[:
VALue]?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):OUTPut:OPTic:POWer:
MAMPlitude:[VALue]?
:<Handle>:SGENeral:PDATa(*):OUTPut:OPTic:POWer:MAMPlitude:[VALue]?
:<Handle>:SGENeral:PPULse(*):OUTPut:OPTic:POWer:MAMPlitude:[VALue]?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):OUTPut:OPTic:POWer:
MAMPlitude:[VALue]?
:<Handle>:SGENeral:PPULse(*):TERMinal(*):OUTPut:OPTic:POWer:
MAMPlitude:[VALue]?
Return Value
Modulation amplitude of the selected connector, terminal, or port.
Examples
:_test:CGR1:MOD1:CONN1:OUTP:OPT:POW:MAMP?
:_test:CGR1:MOD1:CONN1:OUTP:OPT:POW:MAMP:VAL?
might return:
-1.37802405912
:OUTPut:OPTic:POWer:HIGH[:VALue]
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):OUTPut:OPTic:POWer:HIGH:
[VALue] <high power>
:<Handle>:SGENeral:PDATa(*):OUTPut:OPTic:POWer:HIGH:[VALue] <high
power>
:<Handle>:SGENeral:PPULse(*):OUTPut:OPTic:POWer:HIGH:[VALue] <high
power>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):OUTPut:OPTic:POWer:HIGH:
[VALue] <high power>
:<Handle>:SGENeral:PPULse(*):TERMinal(*):OUTPut:OPTic:POWer:HIGH:
[VALue] <high power>
Parameters
<high power>
High power of the selected connector, terminal, or
port.
N O T E
When setting this parameter, the parameter defined by
:OUTPut:OPTic:
POW:UNIT
is used.
This command can also only be used when the
:OUTPut:OPTic:METHod
command is set appropriately.
Examples
:_test:CGR1:MOD1:CONN1:OUTP:OPT:POW:UNIT DBM
:_test:CGR1:MOD1:CONN1:OUTP:OPT:POW:HIGH -0.628620384795
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com