Input Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
251
:INPut:THReshold
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:THReshold <Threshold>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:THReshold <Threshold>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:THReshold
<Threshold>
Parameters
<Threshold>
Threshold value in volts (<NRf>). Specifies the
threshold voltage of an analyzer input connector.
Example
:_test:SGEN:PDAT1:INPut:THR 1
:_test:SGEN1:PDAT1:TERM1:INPut:THR 1
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:THR 1
:INPut:THReshold?
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:THReshold?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:THReshold?
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:THReshold?
Return Value
Returns the current threshold voltage of the specified analyzer input
connector.
Example
:_test:SGEN:PDAT1:INPut:THR?
:_test:SGEN1:PDAT1:TERM1:INPut:THR?
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:THR?
might return.
1.000
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com