Commands in DSR Application Subsystems
Optical Output Commands
286
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Examples
:_test:CGR1:MOD1:CONN1:OUTP:OPT:POW:AVER:VAL?
:_test:CGR1:MOD1:CONN1:OUTP:OPT:POW:AVER?
might return:
0.501187233627E-3
:OUTPut:OPTic:POWer:MAMPlitude[:
VALue]
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):OUTPut:OPTic:POWer:
MAMPlitude[:VALue] <mod. ampl.>
:<Handle>:SGENeral:PDATa(*):OUTPut:OPTic:POWer:MAMPlitude[:VALue]
<mod. ampl.>
:<Handle>:SGENeral:PPULse(*):OUTPut:OPTic:POWer:MAMPlitude[:VALue]
<mod. ampl.>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):OUTPut:OPTic:POWer:
MAMPlitude[:VALue] <mod. ampl.>
:<Handle>:SGENeral:PPULse(*):TERMinal(*):OUTPut:OPTic:POWer:
MAMPlitude[:VALue] <mod. ampl.>
Parameters
<mod. ampl.>
Modulation amplitude of the selected connector,
terminal, or port.
N O T E
When setting this parameter, the parameter defined by
:OUTPut:OPTic:
POW:UNIT
is used.
This command can also only be used when the
:OUTPut:OPTic:METHod
command is set appropriately.
Examples
:_test:CGR1:MOD1:CONN1:OUTP:OPT:METH DBM
:_test:CGR1:MOD1:CONN1:OUTP:OPT:MAMP -1.37802405912