![Agilent Technologies 81250 Скачать руководство пользователя страница 249](http://html.mh-extra.com/html/agilent-technologies/81250/81250_system-programming-manual-and-scpi-reference_2868601249.webp)
Commands in DSR Application Subsystems
Input Parameter Commands
248
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:INPut:TYPE?
Syntax
:<Handle>:SGENeral:
PDATa(*)
:INPut:TYPE?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:INPut:TYPE?
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:INPut:TYPE?
Return Value
Returns the operation mode of the differential input frontend.
Example
:_test:SGEN:PDAT1:INP:TYPE?
:_test:SGEN:PDAT1:TERM1:INP:TYPE?
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:TYPE?
might return:
BAL
:INPut:MODE
Syntax
:<Handle>:SGENeral:
PDATa(*)
:INPut:MODE <DIFF | SNOR | SCOM>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:INPut:MODE <DIFF | SNOR |
SCOM>
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:INPut:MODE <DIFF | SNOR |
SCOM>
Parameters
<DIFF | SNOR | SCOM>
Selects between single-ended operation
and differential operation modes (if supported by the frontend).
• DIFF selects differential frontend operation. Termination is done
according to :INPut:TCONfig, comparator operation is selected
according to :INPut:MODE (this is the reset value).
• SNOR configures the frontend to single-ended operation using the
normal input as shown in the following picture:
Note that :INPut:TCONfig and :INPut:MODE is ignored in this case.
+
-