June 11, 2015, 715004752 Rev. A
Page 112
Figure 4–14: Power-entry module
2. Pinch the sides of the spring-loaded fuse holder, which is below the power entry
module on the rear panel of the detector.
Figure 4–15: Sides of spring-loaded fuse holder
3. With minimum pressure, withdraw the spring-loaded fuse holder.
Figure 4–16: Removing fuse holder
4. Remove and discard both fuses.
5. Make sure that the new fuses are properly rated, and then insert them into the
holder and the holder into the power entry module, gently pushing until the
assembly locks into position.
6. Reconnect the power cord to the power entry module.
Power-entry module
Sides of fuse holder
Fuse holder
Power-entry module
Fuse (2)