Figure 2–13: Data module connections to detector channels A and B
2.6.9 Connecting the detector to a fraction collector
The detector can trigger a fraction collector based on these things:
•
Timed events (see
•
Threshold levels (see
You can connect the fraction collector to one of the detector’s two programmable
switches (SW1 or SW2), and program the timed event, threshold, or ratio at the front
panel. You can also connect the fraction collector to trigger a chart-mark event input
each time a tube is changed at the fraction collector.
1
2
3
4
5
6
7
8
9
10
Analog 1
+
Analog 1
−
Ground
Analog 2
+
Analog 2
−
Switch 1
Switch 1
Ground
Switch 2
Switch 2
2489 connector I
745/745B/746 terminals
Red
Black
1
2
3
4
5
6
7
8
9
10
Analog 1
+
Analog 1
−
Ground
Analog 2
+
Analog 2
−
Switch 1
Switch 1
Ground
Switch 2
Switch 2
2489 connector I
745/745B/746 terminals
Red
Black