B.6 Flow cell specifications
The following table lists the Waters flow cell specifications for the 2489 UV/Vis
detector.
Important:
The Low Dispersion Analytical flow cell is only compatible with the 2489
UV/Vis detector as a component in an ACQUITY Arc system.
Second-order filter
Automatic for wavelengths > 370nm
Wavelength calibration
filter
Erbium filter, used at startup or on demand
Flow cell design
Patented TaperSlit™
a
Nitrogen purge
Purge fitting present on rear of chassis
Path length
10 mm (analytical cell)
Cell volume
16.3 µL (analytical cell)
Pressure limit
6895 kPa (69 bar, 1000 psi) (analytical cell)
Wetted materials
316 stainless steel, fused silica, PEEK, Tefzel
a. US Patent Numbers: US 5,883,721
Table B–6:
Waters TaperSlit flow cell specifications:
Description
Volume (
μ
L)
Pathlength
(mm)
Tubing internal
diameter
(inches)
Pressure
rating
(bar/psi)
Inlet
Outlet
Analytical
16.3
10 0.009
0.009
1000/70
Low dispersion
analytical
16.3
10 0.005
0.005
1000/70
Semi-prep cell
4.4
3
0.040
0.040
1000/70
Microbore cell
4.4
3
0.005
0.005
1000/70
Inert (titanium) cell 16.3
10
0.010
0.010
1000/70
High pressure cell
16.3
10
0.009
0.009
3000/210
Variable path
length flow cell
(VPF)
0.06 to 1.2
0.15 – 3
(factory preset
at 0.5 mm)
0.04
0.04
1000/70
Autopurification cell 6.0
1.0
0.009
0.020
0.040
IN 1
IN 2
OUT
2000/140
Table B–5:
Optical component specifications (continued)
Attribute
Specification