June 11, 2015, 715004752 Rev. A
Page 47
3
Prepare the Detector
After you install the detector, you can set it up and operate it as a stand-alone
instrument or with chromatography data software.
Requirement:
To ensure accurate operation, and before pumping mobile phase or
solvent through the flow cell, be sure to perform the procedures on
3.1 Initializing the detector
Before you power-on the detector, be sure the connector from the rear panel to the
power source is properly installed.
To power-on the detector, press the On/Off switch located on the top left-hand side of
the instrument.
At startup, the detector beeps three times and runs a series of startup diagnostic
tests. If it passes all startup diagnostic tests, the Startup Diagnostics screen appears,
with “OK” following each test.
Figure 3–1:
Startup Diagnostics screen
Table 3–1:
Detector configuration
Configuration
Programming required for operation
The detector is a stand-alone instrument
within a system, or used with any
fluid-handling unit, injector, integrator, or
data system
Program the detector using the front
), except when the
detector is in remote mode.
The detector is controlled by Empower or
MassLynx chromatography data software
Program the detector using the data
system, to control and collect digital
data, following the instructions in the
data system’s online Help.
STARTUP DIAGNOSTICS
TPU
SCI
GPIB
QSPI
OK
OK
OK
OK
ROM
RAM
LCD
CPU
OK
OK
OK
OK