June 11, 2015, 715004752 Rev. A
Page 40
Figure 2–11: e-SAT/IN module channel 1 connection to detector
4. For Channel 2,
•
connect the white wire to pin 4 on I (Analog 2 +);
•
connect the black wire to pin 8 on I (Ground).
Figure 2–12: e-SAT/IN module channel 2 connection to detector
5. Configure the serial port for the e-SAT/IN module as described in the
chromatography data software installation and configuration guide.
The following table summarizes the detector connections to the e-SAT/IN module:
Table 2–6:
Detector connections to a e-SAT/IN module
2489 UV/Vis detector (I)
e-SAT/IN
connector
Pin 1 Analog 1 + (white)
Channel 1 or 2
Pin 3 Ground (black)
Pin 4 Analog 2 + (white)
Channel 1 or 2
Pin 8 Ground (black)
1
2
3
4
5
6
7
8
9
10
Analog 1
+
Analog 1
−
Ground
Analog 2
+
Analog 2
−
Switch 1
Switch 1
Ground
Switch 2
Switch 2
2489 connector I
White
Black
1
2
3
4
5
6
7
8
9
10
Analog 1
+
Analog 1
−
Ground
Analog 2
+
Analog 2
−
Switch 1
Switch 1
Ground
Switch 2
Switch 2
2489 connector I
White
Black
Summary of Contents for 2489
Page 2: ...June 11 2015 715004752 Rev A Page ii...
Page 8: ...June 11 2015 715004752 Rev A Page viii...
Page 14: ...June 11 2015 715004752 Rev A Page xiv...
Page 46: ...June 11 2015 715004752 Rev A Page 46...
Page 124: ...June 11 2015 715004752 Rev A Page 124...