2.2.2 Inspecting the detector
If you discover any damage or discrepancy when you inspect the contents of the
cartons, immediately contact the shipping agent and your local Waters
representative.
Customers in the USA and Canada can report damage and discrepancies to Waters
Technical Service (800 252-4752). Others can phone their local Waters subsidiary or
Waters corporate headquarters in Milford, Massachusetts (USA), or visit
www.waters.com.
For complete information on reporting shipping damages and submitting claims, see
Waters Licenses, Warranties, and Support Services
.
Tip:
Ensure that the instrument serial number, found on the nameplate on the rear
panel or inside the front door of the detector, corresponds to the number on the
instrument integrity certificate.
2.3 Selecting a laboratory site
Install the detector in an area that meets the requirements listed in the table below.
Warning:
For continued protection against fire hazard, replace fuses
only with those of the same type and rating.
Table 2–1:
Installation site requirements
Parameter
Requirement
Operating temperature
range
4 to 40
°
C (39 to 104 °F)
Operating relative
humidity
20 to 80%, non-condensing
Shipping and storage
temperature range
-30 to 60
°
C (-22 to 140 °F)
Shipping and storage
humidity range
20 to 85%, non-condensing
Bench space
12.7 cm (5 inches) clearance at rear
Vibration
Negligible
Static electricity
Negligible