4.6 Replacing fuses
The fuse holder is located on the rear panel of the detector (see figure below). Two
fuses are shipped in place with the detector, rated as indicated in
.
Figure 4–13: 2489 UV/Vis detector rear panel fuse holder
The detector requires two 100 to 240 VAC, 50 to 60 Hz, F 3.15 A, 250 V fast blo, 5
×
20 mm (IEC) fuses.
Suspect a fuse is open or otherwise defective when
•
the detector fails to power-on;
•
the fans do not operate.
To replace the fuses:
Requirement:
Replace both fuses, even when only one is open or otherwise
defective.
1. Power-off the detector, and disconnect the power cord from the power-entry
module.
Warning:
To avoid electric shock, power-off and unplug the instrument before
inspecting the fuses.