1.3 Device Control
46
1.3.13. Settings of Pulse Oxymetry Module
Control of the parameters of the pulse oximetry module (the sweep speed and averaging time of
digital parameters) is carried out using the fields and icons in ECG module window:
To change a parameter, click on the display field of its current value and select the required new
value in the window that appears after clicking.
Default values: sweep speed
– 25 mm/s, averaging time - 16 s.
Decreasing the sweep speed compresses the curve, to view the data over a
longer period of time. Increasing the sweep speed stretches the curve, for
more detailed examination of the curve.
Increasing the averaging time reduces the influence of artifacts on the accu-
racy of the saturation and heart rate parameters (recommended for moving
patients and during surgeries).
Reducing the averaging time allows more quickly to assess the change in
the patient's condition.