Annex 2. Service Menu
190
Annex 2. Service Menu
When the FREEZE mode is activated, it is not possible to enter the techno-
logical menu.
The service menu of the device is used when carrying out adjustment and calibration work with
the monitor, as well as for forced switching on of the drying of the measuring path of the cap-
nometer.
Window of the service menu has the software versions of all modules of the device.
Items of the service menu related to the settings of technological parame-
ters are intended ONLY for technical specialists and service workers during
adjustment or calibration works.
REMEMBER that incorrect actions can lead to unproper settings of the cali-
bration coefficients, which will inevitably affect the measurement accuracy
of the device.
To enter the service menu of the device, press and hold the FREEZE button on the front panel
of the device for 2 seconds. The service menu window appears on the display:
Figure A.2-1. Service menu of the device
Depending on the type of module used for gas analysis, the appearance of
the service menu may differ from Figure A2-1. The differences are described
in the relevant sections of this manual.
In the lower part of the menu window, next to the device ID, there is the version number of the
CPU board / infoport board, and to the right of it - the metrological identifier MetrID (corresponds
to the number of the metrologically significant part of the software), the MAC address of the de-
vice network board and the serial number of the device S/N.