1.3 Device Control
38
For all profiles it is possible to set the patient's weight with an accuracy of 0.1 kg, the minimum
value of the patient's weight is 0.4 kg
To restore the factory settings of the profile, click the corresponding icon:
If the profile parameters were changed during operation by the user and differ from the factory
settings, the display will show the window for returning to factory settings:
If you press the "
Yes
" button, then the set of profile parameters is set in accordance with the
table:
Profiles
Unit
Adult
Pediatric
Neonatal
Patient gender
---
Female
Female
Female
Patient weight
kg
80
35
3.5
Patient height
cm
160
140
55
Age
years
20
4
30 days
Bottom threshold SpO
2
Top threshold SpO
2
%
%
90
---
90
---
80
95
Bottom threshold PR
Top threshold PR
1/min
1/min
50
120
75
160
100
200
Bottom threshold HR
Top threshold HR
1/min
1/min
50
120
75
160
100
200
Bottom threshold Sys.
Top threshold Sys.
mmHg
mmHg
90
160
70
120
40
90
Bottom threshold Dias.
Top threshold Dias.
mmHg
mmHg
50
90
40
70
20
60
Bottom threshold Mean.
Top threshold Mean.
mmHg
mmHg
---
---
---
---
---
---