1.3 Device Control
43
1.3.11. Setting of ECG-channel Parameters
The parameters of the ECG module are controlled using fields and icons located in the cardi-
ometry window:
To change a parameter, click on the field of its current value and select the required new value
in the appeared window.
Setting the sweep speed and averaging time in the ECG channel window
leads to a synchronous change of the sweep speed of the curves and the
averaging time of digital parameters in the PPG and IBP channels.
The ECG scale values correspond to the amplification of the ECG curve: x0.2, x0.5, x1.0, x2.0,
x4.0, x20.0.
A detailed description of the parameters of the ECG module is presented in p. 2.2.
Default values: scale - 1.0 mV; lead - I; ECG filter - filter; cable type - 4 elect.; ECG curves - 1;
mode - Nrm; sweep speed - 25 mm / s; averaging time - 16 s.
When you select an ECG lead that matches the one already displayed on the
screen, the curves of these leads are swapped.
Always check that the set cable type matches the actual one. If there is a
mismatch, some ECG leads may be blocked or the message "Sensor reset"
may be displayed.