1.4 Alarms
62
1.4.8. Alarms of the Capnometry Module
Alarm situation
Message (output
window)
Sound alarm
Light alarm
EtC
О
2
value is less than bottom threshold < 5
%
-
-
Flashing
EtCО
2
value
EtC
О
2
value is less than bottom threshold on
5…10 %
Two short
beeps every 5
s
The same on a
yellow back-
ground
EtC
О
2
value is less than bottom threshold >
10 %
Three short
beeps every 5
s
The same on a
red background
EtC
О
2
value is more than top threshold < 5 %
-
-
Flashing
EtCО
2
value
EtC
О
2
value is more than top threshold on
5…10 %
Two short
beeps every 5
s
The same on a
yellow back-
ground
EtC
О
2
value is more than top threshold > 10
%
Three short
beeps every 5
s
The same on a
red background
FiC
О
2
value is more than top threshold < 5 %
-
-
Flashing
FiCО
2
value
FiC
О
2
value is more than top threshold on
5…10 %
Two short
beeps every 5
s
The same on a
yellow back-
ground
FiC
О
2
value is more than top threshold > 10
%
Three short
beeps every 5
s
The same on a
red background
No breathing
No breathing!
APNEA!* (capno)
Continuous
signal
RSP=0
Thermal stabilization of the measuring cham-
ber elements is in progress
Warming (capno)
-
-
Periodic zero autocalibration is in progress
Zero Calibration
(capno)
-
-
Capnometer pneumatic path occlusion
Line Occlusion
(capno)
-
-
Automatic drying of pneumatic path
Line Purguing
(capno)
-
-
Contamination of the optical chamber
Dirty optics
(capno)
-
-
Measuring chamber is poured (moisture in-
gress)
Off: Poured
Camera (capno)
Two-tone every
5 s
-
Turned off manually by the "CAPNO" button
or icon
Off By User (cap-
no)
-
-
No breath signal at the capnometer input
more than 1 hour
Off: Not Used
(capno)
-
-
Overheating of the measuring chamber ele-
ments
Off: Overheating
(capno)
Two-tone every
5 s
-