6. Troubleshooting
185
6. TROUBLESHOOTING
Trouble symptom
Probable cause
Troubleshooting
“Power” indicator is off when
device is connected to the
mains power
No mains voltage.
The fuse in the mains block has
blown.
Power cord is damaged.
Power system failure.
Check the mains voltage.
Replace the fule.
Replace the power cord.
Contact customer service
Red “Power” indicator
Power system failure
Contact customer service
Short operation time of the
device from the battery
The battery is not fully charged.
BP measurement mode is used too
often.
Reduced battery capacity
Fully charge the battery in accord-
ance with it. 3.4.2.
Increase intervals between blood
pressure measurements.
Carry out training of the battery in
accordance with it. 3.4.3, in case of
failure - replace
Device does not work from
the battery
The battery is fully discharged.
Malfunction of the charger.
Charge the battery, in case of fail-
ure - replace.
Contact customer service
“Battery” indicator is red, con-
tinuously, without blinking
Malfunction of the battery.
Malfunction of the charger.
Replace the battery.
Contact customer service
Spontaneous restart of the
device
Electromagnetic interference when
the device is not grounded
Ensure normal operating condi-
tions (it. 1.2.2)
User settings and internal
clock are not saved
Consequences of long-term stor-
age of the device without switching
on
Turn on the device for several
hours, in case of failure - contact
the service
No sound alarm at all/ for the
particular channel
The alarm volume is set to zero.
Alarm of channel is disabled
Set the required alarm volume.
Enable alarm for this channel
PPG is not displayed, there
are no S
рО
2
and HR readings
The SpO
2
sensor is not installed on
the patient.
Malfunction of SpO
2
sensor.
Install the sensor.
Replace the sensor.
Unstable S
рО
2
and HR read-
ings
Low perfusion at the SpO
2
sensor
application site.
Increased patient mobility.
Malfunction of SpO
2
sensor.
Change the application site or car-
ry out a massage.
Limit the patient's mobility.
Replace the sensor.
ECG waveform and heart rate
readings are not displayed, or
readings are unstable
Poor contact between cable clip
and electrode.
Electrodes are applied incorrectly.
Increased patient mobility.
Electrodes with a long shelf life are
applied.
Check the reliability of the connec-
tion of each clip.
Check the proper application.
Limit patient mobility.
Moisten the electrodes with saline
or contact paste, or use new ones.