2.3 Temperature Monitoring
98
neutral detergent of the "Lotos" type (it is allowed to use other disinfectants of different manu-
facturers, taking into account their methodology applications).
DO NOT ALLOW liquid to get on the sensor connector pins, this can lead to
its failure.
The use of high temperatures and chemical sterilization methods associated
with immersion of the sensor in liquid is not allowed for the disinfection of
the sensors.
DO NOT DISINFECT SENSOR BY IMMERSION IN A DISINFECTING SOLU-
TION!
When disinfecting the sensor cable by pulling it through the disinfectant
sponge, do not apply excessive tensile force to it.
Sterilization of the intracavitary sensor is carried out by full immersion (except for the connect-
or!) in a 6% hydrogen peroxide solution in accordance with the requirements of MU-287-113. It
is allowed to use other sterilizing agents, for example, Deoxon-1 and Dezoxon-4 (Russia) or
Sidex (USA), taking into account the method of their application.
DO NOT immerse the cable connector in a solution or get into it disinfectant
liquids and water.
DO NOT autoclave or sterilize the sensor in a dry oven.
Warranty service for cables that fail as a result of improper operation is not
performed.