3.10 Continuous Non-Invasive Blood Pressure Measurement
168
Message
Description of the situation
Troubleshooting
PWTT:No
ECG signal
One or more ECG electrodes reset, ECG ca-
ble not connected
Connect the ECG cable to the device and
/ or to the patient. Check if the ECG cable
leads are connected to the ECG elec-
trodes.
PWTT:No
PPG signal
SpO
2
sensor is reset or not connected
Connect SpO2 sensor to the device and
(or) patient.
PWTT:No
PWTT data
Available ECG and PPG data do not allow to
calculate PWTT (for example, asystolia or
weak PPG signal)
- check the quality of the ECG electrodes
application, if necessary, replace the elec-
trodes;
- reinstall the SpO2 sensor in a place with
better perfusion of blood circulation.
PWTT: unsta-
ble signal
No clear correlation between ECG and PPG
signals, the results of PWTT measurements
are very different from each other
- check the quality of ECG and PPG sig-
nals, if necessary, reinstall the SpO2 sen-
sor and (or) replace the ECG electrodes
In these cases, dashes will be displayed instead of continuous NIBP readings.
If it is impossible to start extraordinary calibration measurements of blood pressure (the cNIBP
channel is disabled), one of the following messages will be displayed in the cNIBP window:
Message
Description of the situation
Troubleshooting
PWTT off: manual
NIBP
NIBP channel is in manual measurement
mode.
Enable automatic NIBP measurement
mode.
PWTT off: pediat-
ric NIBP
The set profile is
“Pediatric” or “Neonatal”
Set the “Adult” profile
PWTT off by user cNIBP channel is disabled
Turn on the cNIBP channel
In these cases, continuous NIBP readings may be present, but the start of extraordinary blood
pressure measurements is blocked.