1.2 Getting Started
25
1 - Cover lock.
2 -
“ERROR” LED - indicates the absence of paper or loosely closed cover.
3 -
“POWER” LED - indicates power supply.
4 - Paper exit slot.
5 - Cover of the built-in thermal printer.
Figure 1-6. Patient Monitor MPR6-03, side view (the built-in thermal printer cover)
1.2.2. Operating Conditions
Ambient temperature
10…35 °С.
Note. The operating time from the built-in battery is standardized for charging and discharging at tem-
pera
tures from +20 ºС to +25 ºС.
Relative humidity 40 ... 80% (at an air temperature of 25 ° C).
Atmospheric pressure 600 ... 800 mmHg
The choice of the installation site of the device is determined by the convenience of connecting
it to the mains and to the patient, taking into account the fact that an intense external illumina-
tion on the device screen reduces the contrast of the image on it.
The device must be connected to a three-pole socket with a functional protective grounding cir-
cuit (see Fig. 1-7, a). Do not plug the device into a socket without grounding (see Fig. 1-7, b) or
a power extender without grounding (see Fig. 1-7, c).
Figure 1-7. Connecting to the power supply
For the stable operation of the device with electrosurgical devices, it is necessary to connect the
protective grounding wire to the protective earth terminal on its rear panel (see Fig. 1-3, item 5).
In this case it is not allowed to connect the device and the electrosurgical instruments to the
same socket strip (see Fig. 1-7, b) or the power extender (see Fig. 1-7, c).
a (correctly) b (wrong) c (wrong)
1
2
3
4
5