2.1 SpO
2
Monitoring
79
Alarm situation
Message
Sound alarm
Light alarm
S
рО
2
value is more than top threshold by
2…4 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
S
рО
2
value is more than top threshold > 4
%
Three short beeps
every 5 s
The same on a
red background
HR value is less than bottom threshold < 4
%
-
-
Flashing HR
value
HR value is less than bottom threshold by
4…8 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
HR value is less than bottom threshold < 8
%
Three short beeps
every 5 s
The same on a
red background
HR value is more than top threshold < 2 %
-
-
Flashing HR
value
HR value is more than top threshold by
4…8 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
HR value is more than top threshold > 4 %
Three short beeps
every 5 s
The same on a
red background
S
рО
2
sensor is disconnected from the de-
vice
Sensor dis-
conned!*
Two-tone every 5
s
-
SpO
2
sensor is disconnected from patient
Sensor reset!*
Two-tone every 5
s
-
The S
рО
2
sensor is out of order, replace
the sensor
Sensor failure!
Two-tone every 5
s
-
S
рО
2
sensor is incompatible with the
Masimo SET module or is faulty, replace
the sensor
Unrecognized
Sensor!
Two-tone every 5
s
-
The module is out of order, replace the
monitor.
Failure of
operation!
Two-tone every 5
s
-
The module is not installed or out of order,
replace the monitor.
No module
detected!
Two-tone every 5
s
Technical alarms are highlighted with gray fill.
*
– this alarm is triggered if the device has previously recorded readings via the pulse oximetry
channel.
Information messages of pulse oximetry module (Masimo SET)
Message
Description
Connecting to module..
Initialization of the module when the device is turned on
Low PPG signal
Low (weak) pulse filling
Pulse search..
Short-term adaptation to the patient
Too much extern noise!
External interference causes unreliable measurements.
Check the tightness of the sensor to the patient. Check and
Summary of Contents for MPR6-03