1
Getting Familiar with Your TriPlus RSH
MHE Station
48
TriPlus RSH User Guide
Thermo Scientific
The MHE adapter and MHE Adapter Holder dimensions, mass, operating, and
environmental requirements are detailed in
,
.
CAUTION
The MHE Venting Adapter contains a spring-loaded white insert to protect
the needle. Use caution when handling the white insert; the needle can pierce skin.
Note
The MHE Adapter has a threaded gas vent outlet, allowing an optional OD 1/16
inch waste tube to be attached. This optional gas venting tube can be used for applications
requiring hazardous or toxic gases. The pressure setting for the MHE technique is 1.0 to
1.5 bar (about 15 psi). If your system was previously used for static headspace
applications, you may need to adjust your pressure regulator.
Table 15.
MHE Adapter - Dimensions and Mass
Length
Dimension in mm (in.)
Width
25 (1.0)
Height
51 (2.0)
Mass
Dimension in kg (lbs)
MHE Adapter
0.027 (0.06) without vent tubing
Table 16.
MHE Adapter Holder - Dimensions and Mass
Length
Dimension in mm (in.)
Width
35 (1.4)
Depth
150 (5.9)
Height
110 (4.3) with MHE Adapter inserted
Mass
Dimension in kg (lbs)
MHE Adapter
0.285 (0.63) without vent tubing
Table 17.
Operating and Environmental Requirements
Parameter
Requirements
Operating temperature range
4 to 40 °C
Maximum relative humidity
80%, non-condensing
Latitude limitations
3000 meters above sea level