2
TriPlus RSH Basic Informations
Starting the TriPlus RSH
Thermo Scientific
TriPlus RSH User Guide
83
The envelope symbol indicates that a message pending or a
Service Issue
is found.
When powered up, the TriPlus RSH is always in the
User access level.
In this level, the message symbol is visible in the status line, but you have no access rights
to verify or acknowledge the error.
Changing the Access Level to
Extended User
, you can see details of the messages by
selecting each message line and pressing
Enter
. The Pending Message in the example
shows that the reconnected Agitator has not been taught yet.
See the example in
.
Figure 60.
Example of Agitator Missing in Extended User Level
2. To retrieve the message proceed as follows:
a.
Change the Access Level to Extended User.
b. Press the
Options
key to access the pull-up menu for further information.
c.
Select the menu item Pending Message to receive detailed information on the error
message.
A common reason for an error is typically when a
BUS cable
is not connected from one
to another active module. During the boot-up procedure, the presence of all active
modules is verified. If one or more modules are not found, (or it is not possible to check
functions such as
Homing
) an error is recorded. See the example in
.
Summary of Contents for TriPlus RSH