5
Service Menu Item
Installation Item
214
TriPlus RSH User Guide
Thermo Scientific
1. Select the
Cover Plate Type
and the
Vial type
used for the Agitator:
•
Not Specified
— Prohibits the execution of a run.
•
Standard
— Cover without holes.
•
Perforated
— Cover with 6 holes to enable Syringe Needle penetration through the
cover into the vial inserted in the Agitator.
•
Vial Type
— Select the type of vials used for the Agitator.
If the Vial Type has to be changed during routine application, this can be done at any
time by going to the Start Screen, selecting the menu item
Agitator
, scrolling down
to the list item
Vial Type
and selecting another Vial Type.
Teach Module
The term Teaching is used for the definition of the X-, Y-, and Z-Axis positions for each
module. Refer also to
TriPlus RSH Defining Object Positions
in the
TriPlus RSH Hardware
Manual
.
To teach the module
1. Select the wizard item
Teach Module
. A dialog will guide the operator through this
teaching wizard. Upon starting the wizard, the TriPlus RSH automatically turns off the
current to enable a manual movement of the TriPlus RSH Injection Unit to the various
modules configured with the system. See
Note
The difference between
Active
and
Passive
modules are:
• The
Active
modules are recognized automatically by the system when connected to
the Bus and at the next restart of the system.
• The
Passive
modules are not automatically recognized; they have to be added to the
configuration. The passive modules, such as Tray Holder, Wash Station, etc. have to
be added first.