5
Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
215
Figure 157.
Wizard Item Teach Module
2. If the wizard is started without an attached tool, it first moves the TriPlus RSH Injection
Unit to the first slot of the ATC Station or TR Station, picks up the provided tool, and
waits for the next command.
3. The teaching order follows the order of the objects listed as shown in
Setup Modules
(pull-up menu Options). During the teaching process, it is always possible to demand a
fine tuning or to go back to a previous position or module. See the various possibilities
under the
Options
pull-up menu.
4. When the wizard is started, the first module is displayed, an arrow indicates the teaching
position. Follow the instructions given.
As a general description, move the TriPlus RSH Injection Unit manually to the indicated
teaching position, adjust the lower needle guide perfectly within the center of the lunette
ring. Without the magnetic ring attached, use the inner ring for teaching.
5. The L
ower Needle Guide
should be centered on the lunette with the bottom of the
Lower Needle Guide lightly touching its surface.
For modules without a lunette, such as the Agitator, it is necessary to insert a vial (with
the corresponding adapter if required), and use the vial cap surface as the reference for the
Z-Axis value.
The teaching process for the active modules are described in chapter
TriPlus RSH Defining
Object Positions
in the
TriPlus RSH Hardware Manual
.
6. When you are satisfied with the teaching, press
Save
and continue by pressing
Next
to get
to the following module. If in doubt, proceed with the fine tuning.
CAUTION
It is advisable to hold the Y-Axis with one hand and with the other hand hold
the tool body to move the TriPlus RSH Injection Unit in a controlled way to the teaching
position.
Do NOT hold the Needle Guide to move the TriPlus RSH Injection Unit. This part is
mechanically flexible and could lead to misalignment.
Summary of Contents for TriPlus RSH
Page 4: ......
Page 14: ......
Page 206: ......
Page 252: ...5 Service Menu Item Teach Home Position 226 TriPlus RSH User Guide Thermo Scientific ...
Page 258: ......