5
Service Menu Item
Service Menu
190
TriPlus RSH User Guide
Thermo Scientific
Service Menu
The Menu
Service
is only available under the
Extended User Level
. The menu contains the
items
Installation
and
Create New Tray System
. The item
Installation
is a wizard that
guides through the installation procedure covering all points, and controlling the status of the
system. The
Create New Tray System
item will be used during routine operation.
• See
• See
• See
Prerequisites
When the TriPlus RSH is powered on, the
User Access Level
is always active by default.
To enable access to the
Extended User Level
press the two
Menu Buttons
simultaneously, or
press the two characters
A
and
B
simultaneously on the keyboard if you are working with a
Virtual Handheld Controller. For more details, see the section
Figure 134.
Changing Access Level
When the TriPlus RSH is powered up the first time, the
Message Envelope
is shown in the
Menu Screen Status Bar. See the example in
. The reason for this message is that a
TriPlus RSH cannot be operated as long as the motors are not calibrated. The modules cannot
operate if this step has not been completed.
When pending messages are indicating, verify the message. If the message can be explained
with the concept as described above, click it away. For confirmation check all cables from the
active modules are properly connected.