Agitator
Upon selecting the list item
Agitator & Vortexer
, the following parameters are available for
modification. See
Figure 74.
Agitator Parameters
Select
Agitator
, then press
Enter
to edit a parameter. See
Table 31.
Agitator Parameters (Sheet 1 of 2)
Item
Description
Actual Temperature
Reads out the actual temperature of the module; read-only
function.
Max Temperature
Sets the upper temperature limit.
The Maximum Temperature which can be set is 200 °C, but
this parameter is often defined by the specific needs for the
application.
Min Temperature
Sets the minimal temperature limit.
Standby Temperature
The Standby Temperature can be adjusted to the
application’s requirements. Setting the Standby Temperature
using this function does not yet activate the newly-selected
temperature. To activate it, go to the pull-up menu
Options
and select
Activate Standby Temperature
.
The Standby Temperature is activated when the TriPlus
RSH is powered up. In routine use, the Standby
Temperature is activated after a job has been terminated.
The Standby Temperature is typically lower than the
application temperature.
Vial Type Pos. 1
The Vial Type for
Position 1
is displayed. Read-only.
The Vial Type can be set with the task
Setup Modules
.