3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
119
Max Temperature
The Maximum Temperature which can be set is 40 °C, but
this parameter is often defined by your specific needs.
Min Temperature
Sets the minimum temperature
Standby Temperature
The Standby Temperature can be adjusted according to the
application’s requirements. Setting the Standby Temperature
using this function does not yet activate the newly-selected
temperature. To activate it, go to the pull-up menu
Options
and select
Set Standby Temperature
.
The Standby Temperature is activated when the TriPlus RSH is
powered up. In routine use, the Standby Temperature is usually
activated after a job has been terminated.The Standby
Temperature is typically around ambient temperature; this helps to
dry out the internal parts of a Temperature Controlled Drawer,
thus eliminating condensation.
Temp Setpoint
Shows the actual target temperature. Read-only.
CloseSpeed
The speed for closing the drawer can be adapted to a value
between 15 and 100%. (The default value is 80%.)
OpenSpeed
The speed for opening the drawer can be adapted to a value
between 15 and 100%. (The default value is 80%).
Fan Speed
Shows read-only information, and can be used for diagnostic
purposes.
Stdby Temp Control
If switched off, the module will not go to standby
temperature if idle.
Temp. Gain
The coefficient used, if a temperature calibration has been
performed. Possible values are between 0 and 2. Read-only.
Temp. Offset
The Offset temperature used, if a temperature calibration
has been performed. Possible values are between -10 °C and
+10 °C. Read-only.
Table 49.
Stack Module Parameters (Sheet 2 of 3)
Item
Description