• When for technical reasons it is necessary to work on instrument parts which might
involve a potential hazard (moving parts, components under voltage, and so on.) contact
the Thermo Fisher Scientific authorized representative. In general, this type of situation
arises when access to the parts is only possible using a tool. When you perform a
maintenance operation, you must have received proper training to carry out that specific
task.
Electrical Hazards
Every analytical instrument has specific hazards. Be sure to read and comply with the
following pre-cautions. They ensure the safe and long-term use of your TriPlus RSH.
The installation over-voltage category is Level II. The Level II category pertains to equipment
receiving its electrical power from the local level, such as an electrical wall outlet.
Connect the TriPlus RSH only to instruments complying with IEC 61010 safety regulations.
The power line and the connections between the TriPlus RSH and other instruments, used in
the configuration setup of the total analytical system, must maintain good electrical
grounding. Poor grounding represents a danger for the operator, and might seriously affect the
performance of the instrument.
Do not connect the TriPlus RSH to power lines that supply devices of a heavy duty nature,
such as motors, refrigerators and other devices that can generate electrical disturbances.
Use only fuses of the type and current rating specified. Do not use repaired fuses, and do not
short-circuit the fuse holder. The supplied power cord must be inserted into a power outlet
with a protective earth (ground) contact. When using an extension cord, make sure that the
cord also has an earth contact.
If the supplied power cord does not fit the local electrical socket and a replacement or adapter
has to be purchased locally, make sure that only a certified power cord is used. Any power cord
used must be certified by the appropriate local authorities.
Pay attention not to leave any cable connecting the TriPlus RSH and the chromatographic
system, or the power cord close to heated zone, such as the injector or detector heating blocks,
or the GC hot air vents.
Always replace any cable showing signs of damage with another one provided by the
manufacturer. Safety regulations must be respected.
Do not change the external or internal grounding connections. Tampering with or
disconnecting these connections could endanger you and damage the TriPlus RSH. The
instrument is properly grounded in accordance with these regulations when shipped. To
ensure safe operation, you do not must make any changes to the electrical connections or the
instrument's chassis.