• Continue with sample handling but stamp the field
Sample ID
with an error comment
BC not readable
.
• Stop the process of data handling, stop the entire
Sample List
.
This approach of using two scanners confers the advantage that the vial with the barcode label
does not require to be oriented directly towards the scanner. Respecting the minimum length
of 19 mm for the barcode allows the vial be positioned randomly in the and the barcode label
will be successfully interpreted.
Safety Information
The installed laser device is a Class 1 Laser Product.
shows the laser warning label and the location of the laser aperture.
Figure 47.
Laser Warning Label and Laser Aperture
The selected Laser Class 1 for the module Barcode Reader complies with the following
regulations:
• 21 CFR1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50,
dated July 26, 2001.
• EN60825-1:1994 + A1:2002 + A2:2001
• IEC60825-1:1993 + A1:1997 + A2:2001
The software contains a built-in safety time limit such that the laser scanning mechanism
cannot be operated in AIM mode for more than 5 continuous seconds.
WARNING
Class 1 laser devices are not considered to be hazardous when used for their
intended purpose. The following statement is required to comply with international
regulations:
Use of controls, adjustments or performance of procedures other than those specified
herein might result in hazardous laser light exposures.